COLOR CHARACTERISCTICS OF IMAGES, WHICH ARE GAINED WITH LAPAROSCOPS WITH DIFFERENT DIAMETERS, FOR CORRECT MINIINVASIVE DIAGNOSIS

The analysis of the color of laparoscopic and minilaparoscopic images was made using “GretagMacbeth ColorChecker Chart” test. It was established that effectiveness of color reproduction via optic system of laparoscope with minimal diameter (2.0 mm) is 1.75 and 1.47 times less when compared with similar characteristics of optical systems of laparoscopes with grater diameter (10.0 mm and 5.0 mm correspondently). In 83.3% cases of both laparoscopic and minilaparoscopic false positive diagnoses color-deteriorated worsening of informativity of laparoscopic images have been recognized as responsible for the generation of not-corrected diagnostics of abdominal pathology. False-positive minilaparoscopic diagnoses have linked with deteriorated color reproduction two times more frequently when compared with laparoscopic false-positive diagnoses.